| |
NCL Technical Tips
BOD Probes
EVERYBODY wants to blame the probe for their B.O.D. problems, especially for
problems with dilution water blanks. The following procedure will allow you
to check your own probe without sending it out to anybody. Run the following
tests exactly as stated. If the probe fails any of the tests, it needs to be serviced
or replaced. If it passes all of the tests, I give you at least a 95% probability that it is OK.
A. Stability Check:
- Check for air bubbles. Change membrane if necessary.
- Allow AT LEAST two hours after changing a membrane for
the probe to stabilize. Overnight is better.
- Have the probe situated in the air-calibration bottle, and connected
to the meter, and the meter turned on. Remember, “zero”, “red-line”,
and “full-scale” are NOT “ON” positions. Shake or blot
any water droplets off of the probe membrane.
- Allow the system to warm up for at least 30 minutes.
- Shake the probe to remove any droplets of water that have formed on the bottom of the membrane.
- Calibrate the probe by your normal procedure.
- Set a timer for two minutes and observe the readings while the probe
is still in the calibration bottle. Be sure the temperature is constant.
Watch the readings carefully. DO NOT just record the
initial reading and come back 2 minutes later to record that reading.
You need to actually see what happens over the time period.
Following are the possibilities:
If the reading drifts slowly downward, a longer warm up time is required.
If the reading jumps around, there is something wrong with the probe.
If the reading is stable in the air calibration bottle, the sensor is probably good.
(If the readings are stable in the air calibration bottle but not in solution,
the membrane is probably defective.)
B. Zero Oxygen Check: (Response Check)
- Dissolve 3-5 grams of Sodium Sulfite in 300 ml of water. Do this by
stirring slowly so as to avoid cavitation. Slowly pour this solution
into a BOD bottle.
- Calibrate your DO probe as you normally would.
- Place the probe into the “Zero Oxygen” solution bottle
and observe the results.
- Meter should read “0” within two minutes.
C. Temperature Thermistor:
- After calibrating your DO probe, switch the meter to “Temp”,
and record the temperature.
- Using a known good thermometer, measure the temperature and compare
the results. Observe that the temperature reading is stable.
Back to Top
|
- DISTRIBUTOR FOR:
- ALCONOX
- BEL-ART
- COORS
- CORNING
- ERTCO
- DRIERITE
- HANNA
- HF SCIENTIFIC
- KIMBLE
- KONTES
- LABCONCO
- LAMOTTE
- METTLER
- MFS
- MILLIPORE
- NALGENE
- NASCO
- OHAUS
- ORION
- PRECISION
- SPECTRONICS
- THERMOLYNE
- WHATMAN
- WHEATON
- YSI
- WTW
|
|