NCL Technical Tips
High-Performance Ammonia Probes
NOTE: NOT FOR ORIGINAL AMMONIA PROBES
Techniques for Orion’s new High Performance Ammonia Electrode
- None of the old supplies can be used with this probe. This includes filling solution, membranes, membrane caps, ISA buffer, or storage solution.
- ISA buffer – do not use the “blue stuff” for measurements under 20ppm. The new buffer is NCL’s 10N sodium hydroxide (S-41) or Orion’s alkaline solution (O-951011). Use 1ml / 100ml of sample or standard.
- Storage solution – do not use 100 or 1000ppm ammonia standard, use 1ppm ammonia standard (NCL’s A-37C) with 1ml of alkaline solution (or 10N sodium hydroxide) per 100ml.
- Take reading as soon as it is stable or the meter is “Ready”. This probe is fast responding and does not require a 4 minute wait time.
- As always, check and eliminate air bubbles both inside the probe and on the membrane after inserting into solution.
Below are the replacement supplies for the 9512HPBN:
ORION
Cat. No. Description Price
9512HPBNWP |
High Performance ammonia electrode |
$522.00 |
951209 |
High performance fill solution, 60ml |
$ 55.00 |
951214 |
Loose membranes, pk/20 |
$ 98.00 |
951215 |
Pre-assembled outerbody & membrane cap, pk/3 |
$ 50.00 |
951011* |
Low level ISA, Alkaline soln. 475ml |
$ 79.00 |
*The above item is the same as NCL S-41, 10N Sodium Hydroxide |
951211* |
ISA buffer, 475ml, for concentrations above 20 ppm |
$ 75.00 |
*The above item is the same as NCL A-15Y, Ammonia ISA buffer |
Inner Body Check
|
- DISTRIBUTOR FOR:
- ALCONOX
- BEL-ART
- COORS
- CORNING
- ERTCO
- DRIERITE
- HANNA
- HF SCIENTIFIC
- KIMBLE
- KONTES
- LABCONCO
- LAMOTTE
- METTLER
- MFS
- MILLIPORE
- NALGENE
- NASCO
- OHAUS
- ORION
- PRECISION
- SPECTRONICS
- THERMOLYNE
- WHATMAN
- WHEATON
- YSI
- WTW
|
|